Curlin PCA/PCEA Pump

Patient Controlled Analgesia (PCA) with a Clear Lock BoxCurlin Pump

  • PCA therapy is a safe and effective interactive method of pain management.
  • Allows patients to treat their pain by self-administering doses of pain medication within a prescribed time interval.
  • Other infusions using the clear lockbox
    • Continuous IV infusions
    • Subcutaneous infusions
  • Clinical Skills - Elsevier (Mosby's): Continuous Subcutaneous Infusion

Patient Controlled Epidural Analgesia (PCEA) with Yellow Faceplate and Lockbox

24 Hours Technical Support: 1.888.287.5999

 

General Information

Information on Demand Keys (See Reference Card for Keys

Curlin Pump (Click to see larger image)

  • 1 = Date and Time (can ONLY be changed by Clinical Engineering)
  • 2 = Remaining battery power
  • 4 = Holds screens for 6 seconds
  • 5 = Hourly Totals for last 3 hours
  • 6 = First screen of programming
  • 7 = Information on Basal, Bolus, Bolus Interval, Number of Boluses/Hr
  • 0 = Total amount of medication available within the hour

Changing a Basal Rate Or Bolus Amounts

  • Press PAUSE
  • Press YES to RESUME
  • Press NO key to REVIEW program
  • Press YES to accept each line (advances cursor)
  • Enter access code
  • Make change to basal rate
  • Press YES to confirm each line
  • When cursor is on DONE, press YES
  • Press RUN to resume infusion

Review Hourly Totals

  • Press #5 Key, OR
  • Press ‘pink’ Options button (while the pump is running)
  • Scroll down to “Hourly Totals”
  • Press NO to change OFF to ON, then press YES
  • View hourly totals (Boluses Given, Bolus Attempted)
  • 07 = 0700 = 0759
  • Use down arrow to see more totals

Reviewing Hourly Totals (for the last 3 hours, press #5 on keypad)

  • With pump running, press Help/OPTIONS key
  • Move cursor to HRLY TOT
  • Press NO/Change key to change “off” to “on”
  • Press YES to view Hourly Totals
  • Cursor will land on current hour in progress, e.g., 07 is for 0700-0800
  • Use UP arrow to scroll backwards hour by hour
  • Press NO/Change key to exit
  • Press YES to continue

Hanging a New Bag

  • Press PAUSE
  • With cursor on REPEAT RX, Press YES (RESUME” only applies to SAME bag)
    • Pump auto-scrolls through prescription
  • When new bag is in place, press RUN

Pausing and Resuming Same Container

  • Press PAUSE
  • When cursor is on NEW PROGRAM press YES (If Access Code is required, enter it)
  • Enter new values and press YES at each field
  • Press YES when cursor is DONE
  • Press RUN to start infusion

Clinician DOSE:

  • Press HELP/OPTIONS key (pump running)
  • Move cursor to Clin DOSE
  • Press NO/Change Key to change from “OFF” to “ON” (If locked, enter access code and repeat previous step)
  • Press YES to access Clinician Dose Screen.
  • Enter Clinician Dose amount and press YES
  • Press YES to accept and exit the screen
  • Press YES at ACCEPT OPT to exit Options

Changing the Bag

  • Press PAUSE
  • With cursor on REPEAT RX, Press YES (RESUME” only applies to SAME bag)
    • Pump auto-scrolls through prescription
  • When new bag is in place, press RUN

Stopping the Infusion

  • Press PAUSE
  • Press OFF

Documentation

Curlin Pump Documentation

PCA Pump Required Documentation Screenshots

Reference Cards

  • Pocket Card
  • Also available in Central Service (x42215)

On-line resources